Treatment time reduction through parameter optimization in magnetic resonance guided high intensity focused ultrasound treatments

Update Item Information
Publication Type dissertation
School or College College of Science
Department Physics & Astronomy
Author Coon, Joshua
Title Treatment time reduction through parameter optimization in magnetic resonance guided high intensity focused ultrasound treatments
Date 2012-12
Description Magnetic Resonance guided High Intensity Focused Ultrasound (MRgHIFU) treatments are a promising modality for cancer treatments in which a focused beam of ultrasound energy is used to kill tumor tissue. However, obstacles still exist to its widespread clinical implementation, including long treatment times. This research demonstrates reductions in treatment times through intelligent selection of the usercontrollable parameters, including: the focal zone treatment path, focal zone size, focal zone spacing, and whether to treat one or several focal zone locations at any given time. Several treatments using various combinations of these parameters were simulated using a finite difference method to solve the Pennes bio-heat transfer equation for an ultrasonically heated tissue region with a wide range of acoustic, thermal, geometric, and tumor properties. The total treatment time was iteratively optimized using either a heuristic method or routines included in the Matlab software package, with constraints imposed for patient safety and treatment efficacy. The results demonstrate that large reductions in treatment time are possible through the intelligent selection of user-controllable treatment parameters. For the treatment path, treatment times are reduced by as much as an order of magnitude if the focal zones are arranged into stacks along the axial direction and a middle-front-back ordering is followed. For situations where normal tissue heating constraints are less stringent, these focal zones should have high levels of adjacency to further decrease treatment times; however, adjacency should be reduced in some cases where normal tissue constraints are more stringent. Also, the use of smaller, more concentrated focal zones produces shorter treatment times than larger, more diluted focal zones, a result verified in an agar phantom model. Further, focal zones should be packed using only a small amount of overlap in the axial direction and with a small gap in the transverse direction. These studies suggest that all treatment time reductions occur due to selection of parameters that advantageously use mechanisms of decreasing the focal zone size to concentrate the power density, increasing thermal superposition in the tumor, decreasing thermal superposition in the normal tissue, and advantageously using nonlinear rates of thermal dose deposition with increasing temperature.
Type Text
Publisher University of Utah
Subject Cancer; MRIgHIFU; Optimization; Ultrasound
Dissertation Institution University of Utah
Dissertation Name Doctor of Philosophy
Language eng
Rights Management Copyright © Joshua Coon 2012
Format Medium application/pdf
Format Extent 2,518,283 bytes
Identifier etd3/id/2110
ARK ark:/87278/s6862x74
Setname ir_etd
ID 195795
Reference URL https://collections.lib.utah.edu/ark:/87278/s6862x74